NATIONAL
STANDARD
TCVN 9331 :
2012
ISO/TS
22117 : 2010
MICROBIOLOGY
OF FOOD AND ANIMAL FEEDING STUFFS - SPECIFIC REQUIREMENTS AND GUIDANCE FOR
PROFICIENCY TESTING BY INTERLABORATORY COMPARISON
Foreword
TCVN 9331:2012 corresponds to ISO/TS
22117:2010;
TCVN 9331:2012 is developed by the National
Institute for Food Control, requested by the Ministry of Health, appraised by
the Directorate for Standards, Metrology and Quality and issued by the Ministry
of Science and Technology.
Introduction
General requirements for organization of
proficiency testing (PT) schemes of all types are given through ISO/CASCO
(Committee on Conformity Assessment) in ISO/IEC 17043; additionally, general
guidance is available from the International Union of Pure and Applied
Chemistry (IUPAC, see Reference [9]) and the International Laboratory
Accreditation Cooperation (ILAC, see Reference [8]). However, these
recommendations may not be directly applicable to all cases and should be
interpreted specifically for different laboratory sectors where PT schemes are
organized. For this reason, a document is needed to establish the criteria
which a provider (and associated collaborators) of PT schemes shall meet in
order to be recognized as competent to provide PT schemes for microbiological
analysis. This applies particularly to the specific technical requirements
necessary to deal with living microorganisms, such as sample homogeneity and
stability, as well as with the interpretation of presence/absence (detection)
tests which is not covered by an existing document.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Additionally, data from such PT schemes can
be used:
a) to provide information to the
organizations responsible for laboratory acceptance within an official control
framework and to allow continuous monitoring;
b) to aid laboratory accreditation in a
general framework of quality management;
c) to inform those responsible for quality in
the participating laboratories as part of the educative elements of external
quality assessment of trueness (bias).
Information from PT schemes may also be used
for:
1) identification of the possible sources of
errors, particularly the bias component of uncertainty, to improve performance;
2) estimation of measurement uncertainty for
enumeration methods (see ISO/TS 19036[6]) and limits of detection
for presence/absence methods;
3) demonstration of staff competence to
perform a specific microbiological examination;
4) evaluation or validation of a given method
by the study of trueness and precision;
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
6) assignment of a “target” value for an
analyte in a material in order to establish a reference material.
However, these aspects are not specifically
covered in this Technical Specification.
Proficiency testing schemes are therefore
organized to meet certain criteria and the testing programme (frequency, number
of samples, number of repeats, etc.) shall meet the requirements of the type of
method used and commodity analysed, to achieve the level of control desired by
all parties involved.
MICROBIOLOGY OF FOOD
AND ANIMAL FEEDING STUFFS - SPECIFIC REQUIREMENTS AND GUIDANCE FOR PROFICIENCY
TESTING BY INTERLABORATORY COMPARISON
1. Scope
This standard gives requirements and guidance
for the organization of proficiency testing schemes for microbiological
examinations of:
a) food and beverages;
b) animal feeding stuffs;
c) food production environments and food
handling;
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
This standard is also potentially applicable
to the microbiological examination of water where water is either used in food
production or is regarded as a food in national legislation.
This standard relates to the technical
organization and the implementation of proficiency testing schemes, as well as
the statistical treatment of the results of microbiological examinations.
This standard is designed for use with
ISO/IEC 17043 and ISO 13528, and deals only with areas where specific or
additional details are necessary for proficiency testing schemes dealing with
microbiological analyses for the areas specified in the first paragraph.
2. Normative references
The following referenced documents are
indispensable for the application of this document. For dated references, only
the edition cited applies. For undated references, the latest edition of the
referenced document (including any amendments) applies.
TCVN 6910-1 (ISO 5725-1) Accuracy
(trueness and precision) of measurement methods and results - Part 1: General
principles and definitions.
TCVN 6910-5 (ISO 5725-5) Accuracy
(trueness and precision) of measurement methods and results - Part 5:
Alternative methods for the determination of the precision of a standard
measurement method.
TCVN 6404 (ISO 7218) Microbiology of food
and animal feeding stuffs — General requirements and guidance for
microbiological examinations
TCVN 8244-1 (ISO 3534-1) Statistics -
Vocabulary and symbols - Part 1: General statistical terms and terms used in
probability
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
TCVN ISO/IEC 17043:2011 Conformity
assessment — General requirements for proficiency testing
ISO 13528 Statistical methods for use in
proficiency testing by interlaboratory comparisons
3. Terms and
definitions
For the purposes of this document, the terms
and definitions given in TCVN 8244-1 (ISO 3534-1), TCVN 8244-2 (ISO 3534-2),
TCVN 6910-1 (ISO 5725-1), TCVN 6910-5 (ISO 5725-5), ISO 13528, TCVN ISO/IEC
17043 and the following apply.
NOTE 1: Some terms used in the text have
different meanings in microbiology and statistics, e.g. homogeneity,
heterogeneity, test, sample, distribution. The context clarifies whether the
terms refer to microbiological test samples or datasets used for statistical
analysis.
NOTE 2: Some providers of proficiency testing
use the term external quality assessment (EQA) to indicate schemes with broader
application to all areas of operation of a laboratory and a particular
educational remit. The requirements of this standard cover those EQA activities
that meet the definition of proficiency testing.
3.1. target organism
microorganism which is the designated analyte
for a proficiency testing sample.
3.2. background flora
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
3.3. reference strain
microorganism obtained directly from an official
culture collection and defined to at least the genus and species level,
catalogued and described according to its characteristics and preferably
originating from food or water as applicable.
(TCVN 8128-1:2009 (ISO/TS 11133-1:2009)[3],
3.3.2]
3.4. recovery percentage
proportion of the assigned value of the
target organism recovered by the participant.
NOTE 1: The recovery percentage is calculated
by multiplying by 100 the number of recovered colony forming units (cfu) per
volume or per mass.
NOTE 2: The recovery percentage can be
significantly below 100 % when competitive flora and matrix effects are present
in a proficiency testing sample.
4. Scheme design and
purpose
4.1. General
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
4.2. Scheme objectives
The primary objective of any PT scheme is to
provide information to enable laboratories to have confidence in the
reliability of their results.
The detailed requirements for a documented
plan of a PT scheme are covered in TCVN ISO/IEC 17043:2010, 4.4.1.3, and the
plan should also include reference to any relevant legislation. An example of a
plan for a typical microbiology food examination scheme is given in Annex A.
The studies required to establish a new PT
scheme are extensive and shall be clearly defined in the scheme objectives. These
should include, as a minimum, the requirements listed in Clause 5. Requirements
for checking individual rounds of testing, including homogeneity and stability
testing, should also be established in the scheme design and be appropriate for
the scheme objectives.
4.3. Laboratory requirements for schemes
General requirements for appropriate
laboratory facilities to handle all aspects of PT schemes are given in TCVN
ISO/IEC 17043:2010, 4.3.1, and safety requirements are covered in TCVN ISO/IEC
17043:2010, 4.6.2.4.
For microbiology schemes, providers shall have
a documented policy to bring hazards to the attention of participants and
ensure that relevant safety advice is given (see Clause 7). For example, food
microbiology laboratories shall have facilities for dealing with microorganisms
of risk categories 1, 2, and 3, as appropriate (see ISO 7218:2007, 3.2).
4.4. Choice of test matrices
General requirements to document test
matrices in the scheme plan are given in TCVN ISO/IEC 17043:2010, 4.4.1.3, and
choice of the matrices to reflect routine sample types in ISO/IEC 17043:2010,
4.4.2.3.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The description of the test items shall
specify the sample matrix (natural or simulated); whether artificially or
naturally contaminated; the source and country of origin to comply with
international transport regulations; and any method of preservation used, e.g.
freeze-dried, air dried.
4.5. Information on test methods used by the
PT provider
The general requirements for methods to be
used by the PT provider are given in TCVN ISO/IEC 17043:2010, 4.4.1.3.
If the scheme is targeted at one or more
tests specified in or required by legislation, the routine quality control
tests on the scheme samples (e.g. homogeneity and stability) shall be
undertaken in accordance with the methods stipulated in that legislation and
this shall be stated (TCVN ISO/IEC 17043:2010, 4.5.1).
Participants shall be encouraged to use their
routine methods but, where they are undertaking tests in accordance with
legislation, some degree of guidance shall be given, e.g. reference to ISO
methods, legislative texts, or peer-reviewed publications (TCVN ISO/IEC
17043:2010, 4.5.1).
4.6. Statistical design
General requirements for statistical design
are given in TCVN ISO/IEC 17043:2010, 4.4.4.
An outline of the statistical design for PT
schemes for microbiology shall indicate that the statistical tests to be used
are influenced by the level of homogeneity of the test material which, in turn,
is influenced by the random variation in distribution of the microorganisms.
Except for low numbers, a log-normal
distribution is usually expected in quantitative testing data and suitable
statistical analysis methods shall be used for such data [ISO/IEC 17043:2010,
B.3.1.4 d)]. Where low numbers are required in quantitative test items (e.g.
water or beverage examination), a random Poisson model is more applicable, as
the variation in numbers of organisms between different units of material
becomes relatively large and can mask variations in performance.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Semi-quantitative enumeration tests and
qualitative detection tests require different statistical methods to analyse
data and these are discussed further in 8.3 and 8.4.
The scheme plan shall clarify distinctions
between performance testing for methods for detection and those for enumeration
of target microorganisms.
5. Technical
requirements and guidance for sample design and content
5.1. Target organisms level
The target organisms shall be provided at
levels suitable to show that examination methods are fit-for-purpose and to
reflect levels likely to be found in the sample matrices being tested (TCVN
ISO/IEC 17043:2010, 4.4.2.3). Where pathogenic bacteria are the target, the
levels should also take account of and reflect the levels likely to cause
hazard to human health and, if appropriate, any limits specified in
microbiological criteria.
NOTE: The level causing hazard to human
health is not always known with accuracy and depends on the susceptibility of
individuals. The aim of examination for pathogens is to prevent illness, and
also to detect pathogens at a very low level, before those pathogens can grow
to a higher level.
For quantitative (enumeration) methods, the
target level shall be appropriate for the levels routinely found in and any
statutory specifications applicable to the sample matrices used. The target
level should also sometimes be used near the limit of quantification of routine
methods to challenge the performance of the participants across the applicable
range of the method. However, samples should not be dispatched with organism
levels so low that, when using routine dilutions, the expected mean number of
organisms in a sample is fewer than 10 colonies per plate.
For qualitative (detection) methods, the
target organisms shall generally be required to be at a sufficiently low level
to provide a valid challenge to the methodology and to contribute data for
validation exercises to establish or verify limits of detection for individual
participant laboratories.
5.2. Sources, characterization and
traceability of organisms
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Both typical and atypical strains should be
considered and included in the scheme programme to challenge laboratory
performance.
Recognized reference strains from
international collections should be used where they are most suitable for the
scheme purpose; however, laboratory isolates or “wild” strains isolated from
the matrices used by PT schemes are useful to reflect routine situations more
closely. Where these are used, they should be sufficiently characterized
according to the appropriate standard reference methods, to ensure that any
atypical reactions are apparent to the organizers before use.
In all cases, the organisms used in PT scheme
samples should be traceable to the relevant culture collection or to valid
characterization data held by the organizers.
Under certain circumstances, it is not
possible to use reference cultures or materials from internationally recognized
collections or cultured laboratory strains, e.g. for PT schemes for
non-cultivable organisms such as human noroviruses. In such circumstances,
clinical material can be used to contaminate a test matrix artificially, either
through immersion, spraying or, in the case of bivalve shellfish, through
bioaccumulation. The method of artificial contamination should be as close to
the “natural” route of contamination as possible. Extreme care should be used
when manipulating human clinical material, faecal or vomitus samples and these
should be screened for additional pathogens before use. Target viruses should
be fully characterized to strain level by conventional polymerase chain
reaction followed by sequencing.
5.3. Background and competitive flora
The total flora of the samples, either
naturally or artificially contaminated, shall be chosen to assess the ability
of participants to detect and/or enumerate target organisms in the presence of
non-target background flora (typical of the sample matrix) and presumptive
target organisms which, without appropriate confirmation tests, can lead to
false positive results.
Any strains used to simulate background flora
shall meet the requirements of 5.2 for characterization and traceability. In
naturally contaminated samples, the effects of any background flora on the
target organisms shall be determined.
5.4. Matrix selection and effects
All matrices shall be evaluated before use to
check for any effects on the target and background floras, e.g. where the
matrices reduce the recovery of spiked organisms.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Sample matrices used for microbiology PT
schemes are often, but not necessarily, sterilized before use. Where natural,
unsterilized samples are distributed, the organizers shall determine the effect
of the background microflora of the samples.
6. Sample
verification by the provider
6.1. General
General requirements for sample verification
are given in TCVN ISO/IEC 17043 and ISO 13528 (for information, see also
Reference [9]); this clause expands the specific requirements and problems for
homogeneity and stability testing in materials containing living microorganisms.
6.2. Sample homogeneity testing - General
considerations
(See also TCVN ISO/IEC 17043:2010, 4.4.3 and
B.5.)
Proficiency tests may involve the preparation
of a bulk test material, which is then subdivided into individual portions, as
similar as possible to each other, for distribution to participants. Alternatively,
test portions may be individually inoculated for distribution.
Whatever preparation method is used, the test
material shall be assessed for homogeneity, usually prior to but also at the
time of testing for unstable fresh materials.
A homogeneity test should be performed on
each batch of samples, based on relevant statistical principles (TCVN ISO/IEC
17043:2010, 4.4.3.2 and B.5). Such tests are given in ISO 13528 or, as an
alternative, Annex B.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
A test material which is less than
sufficiently homogenous may still be used in a proficiency test round (TCVN
ISO/IEC 17043:2010, 4.4.3.1 Note 3), provided suitable statistical principles
are used to take account of the greater variance between samples (see ISO
13528). A statistical plan for heterogeneous materials, including replicate
analysis of several samples [(see TCVN 6910-5 ISO 5725-5)], should be used to minimize
the effects of lack of homogeneity on the evaluation of participant
performance.
6.3. Homogeneity testing for quantitative
(enumeration) samples
General requirements and procedures for
testing homogeneity of quantitative proficiency test materials are given in
TCVN ISO/IEC 17043:2010, 4.4.3 and B.5 and ISO 13528.
Materials that show between-unit variation
large enough to affect the assessment of laboratory performance significantly
should not be used in interlaboratory studies, unless special requirements and
methods of data analysis apply, e.g. low numbers of microorganisms in drinking
water and other samples.
The criterion for “sufficiently homogenous”
is defined by the requirements of the interlaboratory comparison (see ISO 13528
and Annex B). However, in general, a material for which the between-unit
standard deviation (on the appropriately transformed scale) is ≤ 0,3 sp, where sp is the target
standard deviation used to assess the performance of laboratories, is
considered sufficiently homogenous (see ISO 13528).
Any alternative homogeneity test should meet
the following criteria (reproduced from Reference [9]):
a) the probability of rejecting a
sufficiently homogenous test material should be ≤ 5 %;
b) the probability of rejecting a test
material where between-unit variation is 1,5 sp, in which sp is the acceptable between-laboratory
variation (expressed as a target standard deviation), is W 80 %.
An 80 % probability of rejecting a material
where between-unit variation is 1,5 sp is based on simulation studies of the duplicate analysis
of 10 test units using a method with an analytical standard deviation of 0,5 sp (i.e. 0,125 log
units) and a critical value for T2 (see Annex B) that meets criterion a) in the
previous paragraph. It represents what is achievable with a reasonable amount
of analytical effort.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The analytical (repeatability) variance
should be estimated from replicate analyses of the initial suspensions obtained
from test portions (References [15][16]). This analytical variance can also be
calculated from the number of counted colonies and the precision of analytical
materials in use (Reference [10]).
In microbiology, the between-unit variance
shall be estimated under repeatability conditions (in one run). If that is
impossible, the between-unit variance includes the within-laboratory
reproducibility and can perhaps lead to the false rejection of a satisfactory
material.
When the number of counted colonies is
sufficiently high (more than 35 to 40 colonies per plate), the analytical
standard deviation, san, generally satisfies:

where sp is the target standard deviation, and the test for
sufficient homogeneity proposed in Reference [11] should be used (see Annex B).
If the number of counted colonies is low (fewer than 35 to 40 colonies per
plate), the T1 − T2 test is recommended
(see Annex B).
When replicated test units are provided to
participating laboratories, the between-unit variability obtained by
participants should be examined by the provider to assess the homogeneity of
the material. Although this variability includes within-laboratory
reproducibility, the higher number of participating laboratories increases the
statistical power of the analysis and can be a good indicator for successive
rounds.
When the number of counted colonies is low
(say fewer than 20), the analytical (repeatability) variance is high. In that
case, the provider should recommend that participating laboratories replicate
enumerations of test portions to satisfy the condition (see ISO 13528):

where:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
sp is the target standard deviation;
n is the number of replicates.
If that is impossible, the laboratory
performance should be assessed cautiously.
Where the PT material contains low numbers of
cfu (say fewer than 20), the between unit (i.e. between replicate samples)
variation shall be measured to demonstrate that it does not exceed random
(Poisson) variation in order to provide a meaningful assessment. The index of
dispersion test on n samples (where n is a minimum of 10) should not exceed the
value at 0,05 probability.
6.4. Homogeneity testing for qualitative
methods
Similar principles apply to homogeneity
testing for qualitative (presence/absence) methods, but special consideration
is required where the level of spike is low (see 8.4).
The homogeneity of qualitative samples may be
tested by enumerating the spike. It may also be possible to determine the
contamination level by using a most probable number (MPN) technique. If
enumeration is possible, the homogeneity tests detailed in 6.3 may be used,
depending on how the samples are artificially contaminated and on how the spike
is enumerated.
6.5. Stability testing by the provider
6.5.1. General
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
If the same type of sample with the same test
strain(s) is always used, it is sufficient only to perform a verification (e.g.
by checking after preparation and at the date of the study) on subsequent
samples. The provider should also examine results obtained by the participating
laboratories to check the stability of the material during the period allowed
for the study.
6.5.2. Stability during storage conditions
For stable samples which are always stored at
low temperatures (e.g. −70 °C, −20 °C, +5 °C), the stability shall be
determined by checking the level of the analyte and the homogeneity of each
batch at regular intervals during storage. The minimum period for stability
testing should be the time between preparation of the materials and the
specified date or time period of analysis.
Frequency of testing depends on the
information already available for the batch of samples and the total period of
time over which stability information is required. If the total storage time
is, for instance, only two weeks, it may be necessary to test every two days,
but if storage is required for one year, it may be sufficient to test monthly. For
large batches of samples, a minimum of three samples should be tested on each
occasion in order to show ongoing stability across the whole batch (ISO
13528:2005, Annex B). Whatever frequency of testing is used, this shall be
justified and validated as acceptable by the scheme organizers.
6.5.3. Stability during transport conditions
In addition to information on stability
during storage conditions gathered during validation studies for a new scheme,
it is also important to test the effect of “abuse conditions” on the samples,
e.g. long transport times at elevated temperatures (see TCVN ISO/IEC
17043:2010, 4.6.3.2).
A stability test using different temperatures
to reflect “worst case” transport conditions and maximum expected delays, shall
be performed initially to test the effect of such abuse on test samples. For
example, samples of one batch are stored at the specified storage temperature
(e.g. −20 °C), but also at +5 °C, +15 °C and +25 °C. Every day, five samples
held at each storage temperature are analysed for a total period of one or two
weeks.
The design of such stability experiments is
variable but should be appropriate to obtain information on the effect of
different storage temperatures on the samples and to establish any upper
temperature limit for receipt by the laboratory. The information obtained can
be used to choose the optimal distribution conditions for the scheme samples,
e.g. whether it is necessary to cool the samples during transport using dry ice
or ice packs, or whether ambient distribution is acceptable.
For temperature-critical distributions
requiring controlled refrigeration with strict sample acceptance criteria, such
as schemes in support of legislative testing for E.coli in bivalve shellfish,
inclusion of individual temperature loggers in each sample box to record the
sample temperature in transit is recommended. Checking the temperature data may
enable the scheme provider to explain anomalous results returned by
participants.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
7.1. General
Requirements for general sample handling are
detailed in TCVN ISO/IEC 17043:2010, 4.6.1 and 4.6.2, and only additional
information relevant to microbiological samples is given in this clause.
7.2. Instructions to participants
For each study, each participating laboratory
shall receive a clear set of instructions covering:
a) storage conditions for samples of all
types, particularly information on the storage temperature, which should also
appear on the outside of the transport packaging;
b) maximum temperature of the samples on
receipt at the participant laboratory, if appropriate;
c) instructions on how to handle the samples
— if reconstitution, dilution or other processing of the samples is required,
this should be described clearly for each set and type of samples;
d) appropriate safety data sheets, which
should be included with each distribution (an example of the detail required is
given in Annex D);
e) other supplementary instructions, such as:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
2) the method(s) of examination (prescribed
or participant choice, as required);
3) how to report the results to the
organizers, particularly the units of measurement;
4) the organizers may request details on
materials, methods, incubation conditions, etc., in report proformas - if this
is the case, instructions for completing such proformas should be provided.
8. Performance
evaluations
8.1. General
Wherever possible, the statistical principles
used to evaluate performance in PT schemes should be based on those given in
standards, such as ISO/IEC 17043:2010, 4.7 and ISO 13528 (for information, see
also Reference [9]), although microbiology PT schemes may adopt procedures
which differ from those commonly used in other sectors if they are appropriate
to their particular schemes.
8.2. Preliminary considerations
Proficiency testing involves the regular
distribution of test materials to participating laboratories for them to
examine the test materials for specific measurands (in most cases
microorganisms). The results of examination are then compared against those of
other participants.. Proficiency testing therefore provides an independent
means of testing and comparing individual performance.
Ongoing satisfactory performance in
proficiency test rounds can provide reassurance to participants of their laboratory
processes, including methods of examination, analyst training, equipment,
reagents, quality control procedures, interpretation of results, and reporting
techniques.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
There are many different ways to interpret
data from proficiency testing but methods for interpretation shall be
objective.
The majority of participants taking part in
microbiological PT schemes are not familiar with statistics and shall have
confidence that the procedures used by the PT scheme organizers are sound.
The following considerations about the statistical
principles applied shall be addressed:
a) validity;
b) explanatory information for participants;
c) reasons for selection;
d) consistency.
Clear and unbiased information shall be
provided to participants to allow self-assessment and interpretation of results
and to maximize the benefit from participation.
8.3. Quantitative
methods
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The statistical design for PT schemes for
microbiological enumeration methods is influenced by the level of homogeneity
of the test material, which in turn is influenced by the random variation in
distribution of organisms. In addition, there are likely to be substantial
differences between the participants in the precision required or expected of a
test.
The choice of statistical method shall take
into account the factors outlined in 8.1 and 8.2, together with other
considerations such as the number of participants in a particular scheme. Parameters
that contribute to deciding which statistical tests are to be used for
analysing results should be stated. For example, specialized schemes may have
fewer than 30 participants; results from 30 participants may be analysed in a
different way from larger schemes with, for example, more than 100
participants.
The chosen method of statistical analysis
shall be appropriate for not only the number of participants undertaking an
examination but also the method used. For example, results for an examination
using colony count methods are analysed in a different way to those where
determination of the MPN is required. This is because the inherent variation in
the MPN method tends to be greater than that for colony count methods. Indeed,
different statistical parameters may be needed to assess participants'
performance for different tests within a single scheme (see ISO/IEC 17043:2010,
4.5.2). This shall be stated in the scheme design documents and generally
accepted microbiological criteria shall be referenced in the scheme plan.
Suitable statistical analyses and allocation
of scores to participants are covered in TCVN ISO/IEC 17043:2010, Annex B.
8.3.2. Distribution of data
When an enumeration test is repeated several
times, under repeatability conditions, the frequency distribution of the
results would be expected to form a bell-shaped curve, called a normal
distribution.
Microbiological counts usually follow a
log-normal distribution and the data are converted to logarithm to the base 10
values to produce a normal distribution curve. However, for low bacterial
numbers, actual counts may be used, or a square-root transformation may be
applied.

KEY:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
m
mean
s
standard deviation
Figure 1 - Diagram of
a normal distribution
Although it is the same material under test,
the results are not all identical, as numerous small, independent variations
are expected to occur during the different manipulations involved in performing
the tests.
In proficiency testing, the tests are not
usually performed under repeatability, or even reproducibility conditions. Tests
are performed by different analysts, at different test locations and at
different times, using a variety of equipment, media, reagents and analytical
methods and this results in further variability which could be described as
“super-reproducibility” or “over-reproducibility” conditions because the term
reproducibility is usually used in the context of a single method only.
Despite such variations in test conditions, a
(log-)normal distribution of the results is usually observed and the principles
of statistics appropriate to (log-)normal distributions should be used to interpret
the data, provided the distribution is roughly symmetrical and unimodal.
If the distribution of data does not appear
(log-)normal, the possible reasons should be assessed and the data interpreted
accordingly using other suitable tests.
8.3.3. Determining the assigned value
The purpose of proficiency testing is to
assess how proficient participants are in achieving the “correct” result. However,
in many proficiency testing schemes, it is not possible to know the “correct”
result, as numerous analysts may all examine the same test material and all
return a slightly different result.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The most usual method for microbiological PT
schemes is consensus values from participant laboratories. The assigned value
is determined from the robust mean or the median of the results of all participants
The use of robust methods is intended to minimize the influence of outliers, so
that such results need not be excluded from the data because “true” outliers
may be difficult to identify.
The assigned value has a higher uncertainty
than with other methods, but this is taken into account when assessing
performance. The assigned value is deemed to be fair because all participants'
results contribute to the calculation
If a low overall median is produced where
assigned values are set from participant consensus (e.g. because a large number
of participants had difficulty isolating or identifying a particular organism),
the scheme organizers should comment accordingly, so that the performance of
participants whose results were not affected is correctly judged.
8.3.4. Uncertainty of the assigned value
The assigned value represents the best
estimate of the “true” value. It has a standard uncertainty indicating the
level of confidence in this estimate. If the standard uncertainty of the
assigned value is too large in comparison to the standard deviation of the test
round, then some participants receive action and warning evaluations, not
because of their performance but instead due to the large uncertainty in the
assigned value.
Criteria for the acceptability of an assigned
value in terms of its uncertainty should therefore be established. A number of
methods for estimating this uncertainty and determining acceptability are
available and are described in ISO 13528.
8.3.5. Methods of assessing performance
The organizer of the PT scheme shall
determine the assigned value and assess by how much the result(s) from each
individual participant deviate(s) from that assigned value compared with the
results from all other participants. Thus participant performance is judged not
against the “correct” result, but against statistical estimates of the
“correct” result derived from all of the submitted data. The larger the number
of data, the more accurate such statistical estimates are likely to be.
Common methods of assessing performance and
allocating scores are detailed in TCVN ISO/IEC 17043:2010, B.3 and ISO 13528.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
8.3.6.1. General
A common and widely accepted method used in
proficiency testing is the z-score system, as this is relatively easy to calculate
and interpret (see TCVN ISO/IEC 17043:2010, B.3). The z-score indicates how
many standard deviations away from the mean a given value lies, e.g. a z-score
of 2 represents a value which is 2s,
where s is standard
deviation, from the mean.
As depicted in Figure 1, data with a
standardized normal distribution have 95 % of values within 2 σ of the mean and
99,7 % of values within 3 σ. Results with a z-score greater than 2 are
therefore considered questionable because only 5 % of correct measurements are
expected to be that different from the assigned value. Results with a z-score
over 3 are considered unsatisfactory because only 0,3 % of correct measurements
are expected to be that different from the assigned value (see TCVN ISO/IEC
17043:2010, B.4).
8.3.6.2. The target standard deviation for
z-score calculations
The z-score calculation uses a target value
for standard deviation. This target standard deviation defines the scale of
acceptable variation among laboratories for each particular test. The same
target standard deviation should be used over successive rounds of the
proficiency test so that scores may be compared from round to round. There are
a number of methods for establishing the target standard deviation, detailed in
Reference [9], TCVN ISO/IEC 17043:2010, B.3 and ISO 13528.
8.3.6.3, Multiple results in z-score systems
Differences between participants' results
arise from between-laboratory variation and also from within-laboratory
variation. Within-laboratory variation or intra-laboratory variance is the
variation between measurements made by the same laboratory on the same sample
and is an inherent feature of microbiological examinations. Within-laboratory
variation is measured by the repeatability variance.
When a participant laboratory reports
multiple results for a single test material, this may potentially bias the
remaining data from other participants. For example, if 10 analysts from the
same laboratory all tested the same sample, which had been incorrectly diluted
initially, all 10 results would be incorrect. This number of incorrect results
becomes a subset within the bulk data, and may bias all the other results. To
avoid such bias, only one reported result per laboratory should be included in
the overall analysis of data for a distribution.
When a participant laboratory has obtained
multiple results from a single PT sample, these results shall be reported
separately and not as a mean value. Where only one result per participant is
permitted by the scheme organizers, the individual result to be reported shall
be chosen before the examination is undertaken and results are known.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
8.3.7.1. General
Although z-scores are very commonly used for
evaluation of PT scheme results from enumeration methods, other methods of
scoring may be appropriate for particular schemes.
For example, with samples where low counts
are sought (such as drinking water) the statistical assessment can be based on
a model which predicts random variation. Thus “low” or “high” tail-end counts
are defined using the Poisson formula. Low or high results can occur
occasionally by chance in any laboratory, but an accumulation of tail-end
results indicates poor performance. The advantages and disadvantages of a model
versus a percentile approach to statistical assessment are discussed in
Reference [17].
With schemes where high counts are expected,
the 0,5 log10 rule or the percentile approach can be used, but the median
absolute deviation method is required for schemes with low numbers of
participants. These are briefly outlined in 8.3.7.2 to 8.3.7.4.
8.3.7.2. Using the 0,5 log10 rule
A scoring system based on the 0,5 log10
rule can be used for colony counts (adapted from Reference [13]). In summary,
the 95 % confidence intervals around a mean colony count are generally not more
than ±0,5 log10 cfu. Internal quality control procedures for
microbiology laboratories commonly require replicate counts to show agreement
to not more than 0,5 log10 units to demonstrate good control. This
is applied to participants' results such that all results within ±0,5 log10
units of the participants' median are considered as acceptable and are
allocated the maximum score. This rule allows participants' scores to improve
over time if the overall quality of participants' enumerations also improves.
The 0,5 log10 rule is based on
microbiological criteria but is also statistically valid because if the
expected count on a plate is 10 colonies, and organisms are randomly
distributed, then 95 % of results should show between 3 and 17 colonies. On a
decimal logarithm scale, the expected median count is 1, the lower limit is
0,47 and the upper limit is 1,23, i.e. the lower and upper limits are within
0,5 log10 units. Therefore, a result of within ±0,5 log10
units of the expected value should be deemed acceptable.
8.3.7.3. Using percentiles
Percentiles can be used to identify outlying
counts for enumerations when W 50 participants undertake an enumeration using a
colony count procedure. This entails calculation of the 5th, 10th, 90th, and
95th percentiles of the distribution of participants' results (C5, C10, C90,
and C95 respectively). C5 and C10 should be rounded down to the nearest 0,05
log10 unit (e.g. 2,23 rounds to 2,20), whereas C90 and C95 should be rounded up
(e.g. 3,36 rounds to 3,40). An example of how scores might be allocated (e.g.
for aerobic colony counts) is outlined below:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
- Results between C5 and C10 or C90 and C95: Score
= 1
- Results below C5 or above C95: Score
= 0
Application of the 0,5 log10 rule
may extend the acceptable range and therefore upgrade the scores allocated for
some results in C5, C10, C90 and C95.
The percentile method is robust and does not
depend on the actual distribution of decimal logarithm counts being normal. It
enables a clear interpretation of the performance assessment.
If the number of participants in an
established PT scheme falls to fewer than 50 laboratories or the number of
participants reporting enumeration results for a particular parameter falls to
less than 50, then median absolute deviation (MAD) values (see 8.3.7.4) should
be used instead of percentiles.
8.3.7.4. Using median absolute deviation from
the median values
The MAD method is used to identify outlying
counts when fewer than 50 participants undertake an enumeration. Percentiles
should not be used because fewer than 50 results provide insufficient data to
calculate valid values for C5, C10, C90 and C95. MAD values provide a robust
method for calculating the acceptable range when assessing participants'
results and allocating scores. The analysis requires calculation of the median
difference from the median for every result which is then multiplied by 1,4826
to get a robust estimate of the standard deviation (MAD value), sMAD.
An example of how scores might be allocated
using MAD values is outlined below:
- Results within participants' median ± 2sMAD: Score = 2
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
- Results outside ± 2,58sMAD: Score = 0
As with percentiles, the lower limits should
be rounded down to the nearest 0,05 log10 value and the upper limits up to the
nearest 0,05 log10 value.
Note that, unless the 0,5 log10
rule is applied, approximately 5 % of laboratories should be outside the ± 2sMAD range and 1 %
outside the ± 2,58sMAD range (assuming
normality of the decimal logarithm counts without extreme outliers).
The MAD method should be used for new PT
schemes where there are fewer than 100 participants.
8.3.7.5. Special considerations for most
probable number methods
The test method used to determine an MPN
value has greater inherent variability than colony count methods and is
therefore often regarded as only semi-quantitative. However, it is sometimes
required for the detection and estimation of levels when low levels of
microorganisms are expected, especially when the microorganisms may be stressed
(e.g. as a result of processing or freezing). Also, MPN methods are stipulated
in legislation or criteria for the microbiological examination of certain
products such as dairy products and live bivalve molluscs and other shellfish.
Any method for assessing participants'
results for determining MPN values should allow for the inherent variability of
the MPN, and assume that the sample is well mixed prior to testing.
For the three-by-five tube method, the
standard deviation of a log10 MPN result is approximately 0,24, provided
results do not show “extreme” tube combinations, e.g. tube combinations of 3,
0, 0 to 5, 5, and 2 [see TCVN 6404 (ISO 7218)].
For the three-by-three tube method, the
standard deviation of the log10 MPN result is approximately 0,32,
provided results do not show “extreme” tube combinations, e.g. tube
combinations of 2, 0, 0 to 3, 3, and 1.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
However, in practice there is some
between-laboratory variability. Analysing a number of sets of data has shown
this to inflate the variance by about 2,5-fold (and hence the standard
deviation by about 1,58-fold).
Therefore the limits of acceptability for
participants' results for MPN determinations should be raised to ±3s and ±5s (see Table 1).
Table 1 - Limits of
acceptability
Limit of
acceptability
Three-by-three
method
Three-by-five
method
± 3s
± 0,96 log10
(9,1-fold)
± 0,72 log10
(5,2-fold)
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
± 1,60 log10
(39,8-fold)
± 1,20 log10
(15,8-fold)
The 0,5 log10 rule should never
need to be applied to MPN results due to the inherent method variability.
It is also possible, for the MPN method, to
check that the tube combinations and dilutions reported are consistent with the
MPN reported using tables [see TCVN 6404 (ISO 7218)].
If the PT scheme or the legislation on which
it is based requires MPN values to be determined in duplicate, the results can
be compared and if the tube combinations are credible, then the difference
between the two results should not differ, in terms of decimal logarithm units,
by more than 2,58 x
x 0,24 = 0,88 for the
three-by-five tube method and 2,58 x
x 0,32 = 1,17 for
the three-by-three tube method.
If two distributions with two replicates per
distribution are compared, then the mean of the two should not differ, in terms
of decimal logarithm units, by more than 2,58 × 0,24 = 0,62 for the
three-by-five tube method and 2,58 × 0,32 = 0,83 for the three-by-three tube
method.
8.3.8. Long-term performance assessment
8.3.8.1. General
Performance assessment in proficiency testing
schemes is generally confined to assessment of results from single rounds, but
there are instances where assessment in the longer term may be beneficial.
Whilst this generally applies to external quality assessment schemes, some
guidance is given for the sake of completeness.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
“Low” and “high” counts should be defined
according to objective rules (e.g. Poisson model-based definition for low
counts, percentiles or other methods for high counts), then used to determine
those laboratories reporting such results more frequently over time than could
be expected by chance.
Scheme organizers should encourage
participants to exercise their own professional judgment to assess information
supplied in reports, thereby self-assessing their performance. Organizers may
suggest to their participants various ways of undertaking laboratory
self-assessment but they should not dictate criteria for self-assessment by an
individual laboratory; rather these should be set by each individual
participating laboratory, based on what it believes to be microbiologically
significant in the context of its own routine work and client requirements.
8.3.8.2. Low count assessments
f chance is the only factor involved, the
“tail-end” counts should be distributed at random. The results may be
scrutinized to determine the scatter of tail-end results, between laboratories
over a series of samples (e.g. using Cochran's Q-test).
If they are not distributed at random, a
second stage analysis may be performed to determine the expected distribution
of tail-end counts, amongst those laboratories reporting them, if they were
simply due to natural variation between samples and not to laboratory performance
effect. Then, contrasting those expected numbers of laboratories with the
number actually observed highlights those laboratories that may have
experienced problems. An example of performance assessment for samples
containing low numbers is given in Table 2.
Table 2 - Observed
and expected numbers of sets of results assuming random distribution of low
results (from a distribution of low levels of Clostridium perfringens in
drinking water samples, where variation in numbers between test items may
necessarily exceed laboratory performance variation)
No. of “lows”
Observed
Expected
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
58
58
1
32
48,56
2
18
16,94
3
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
3,15
4
3
0,33
5
2
0,02
Total
127
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
One or two low results could have been due to
chance; three were possibly not due to chance; four or more were unlikely to be
due to chance and procedures should be checked.
8.3.8.3. High count assessments
Assessment of higher counts relies on similar
principles, but less variation in participants' results due to variation in
sample content is expected.
Long-term performance with, for example, the
percentile method of evaluation, can be assessed over 12 samples, allowing a
maximum score of 24 points. Participants are set a performance target of at
least 70 %. If the 0,5 log10 rule is discounted and an assumption is
made that all participants are capable of delivering equivalent performance,
then using multinomial theory, it can be shown that the probability of a
participant obtaining a cumulative score that is less than 70 % of the maximum
possible score is 5,2 %. This means that under these circumstances,
approximately 1 in 20 participants may be identified incorrectly as “poor
performers”. In reality performance is not equivalent and some laboratories do
experience difficulties with the examinations that they undertake, so the
probability of a satisfactory performance being incorrectly identified as
“poor” is much lower than this. Furthermore, the use of the 0,5 log10 rule
reduces this probability to less than 0,1 %.
A practical example of long-term performance
assessment using spreadsheets is shown in Annex C.
8.4. Assessment of
qualitative methods
8.4.1. General
For interlaboratory comparison studies in
which one or more qualitative methods are used, the results are in fact black
or white, yes or no, detected or not detected.
Methods of statistical analysis for this type
of result are limited, but various options have been proposed, such as LOD50,
accordance or concordance assessments and percentage accuracy, and the optimal
approach is still under consideration.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
A simple method for self-assessment by
participant laboratories is to record the number of positive and negative
results they have found, together with the number of positive and negative
results which were expected. This information should be linked with records of
the levels of target organisms in the samples to assess performance and also
provide ongoing data on the limit of detection of the method in individual
laboratories.
As the result of a presence/absence test is a
yes or no, many samples need to be tested in order to assess the performance of
the individual laboratories in a PT scheme. Each participant should test at
least 18 samples in total. These 18 samples consist of six replicates of three
different levels of contamination of the samples. The three levels are:
negative (check on the occurrence of false positive results due, for example,
to cross contamination); low level [meaning samples contaminated at or slightly
above the detection limit for the method used, which ideally should be at the
level where 50 % of the samples are found positive; and 50 % negative (LOD50)]
and high level (this level should be 10 times higher than the low level and
represents the level at which all samples tested should be found positive).
The interpretation of the data is simple:
a) for the negatives: all samples should be
found negative;
b) for the high level: all the samples should
be found positive;
c) for the low level: it can be calculated
(see Table 3) using the binomial distribution and the percentage of samples
found positive (can be obtained from a reference value from the organizer or as
a best estimate from the results of all participants) at a 95 % confidence
level.
Table 3 - Chance of
finding a certain number of positives out of six samples tested as a function
of the average percentage of positive samples (binomial distribution)
Number of positives
out of six samples
Average percentage
of positives
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
20%
30%
40%
50%
60%
70%
80%
90%
0
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
26,2 %
11,8 %
4,7 %
1,6 %
0,4 %
0,1 %
0,0 %
0,0 %
1
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
39,3 %
30,3 %
18,7 %
9,4 %
3,7 %
1,0 %
0,2 %
0,0 %
2
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
24,6 %
32,4 %
31,1 %
23,4 %
13,8 %
6,0 %
1,5 %
0,1 %
3
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
8,2 %
18,5 %
27,6 %
31,3 %
27,6 %
18,5 %
8,2 %
1,5 %
4
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
1,5 %
6,0 %
13,8 %
23,4 %
31,1 %
32,4 %
24,6 %
9,8 %
5
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,2 %
1,0 %
3,7 %
9,4 %
18,7 %
30,3 %
39,3 %
35,4 %
6
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,0 %
0,1 %
0,4 %
1,6 %
4,7 %
11,8 %
26,2 %
53,1 %
Example: To use Table 3, first the average
percentage of positives has to be known. Here it is assumed to be 30 %, meaning
that about one out of three samples contains the target organism. As only 30 %
of the samples contain the target organism, it is likely that some of the
samples might not contain the target when, for example, six samples are tested.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Another reasoning is that the chance of
finding six out of six positives is only 0,1 %. This is very unlikely to happen
purely by chance. As the uncertainty is set at a maximum of 5 % (100 % – 95 %
confidence level), this falls within this limit. The same occurs for the situation
when five or six samples out of a total of six are found positive. The sum of
these chances is 1,1 % which is still below the 5 % limit. Only when the
situation where four out of six are positives (6 %) is added does the
uncertainty exceed the 5 % limit. As the maximum is set at 5 %, the situation
of five or six positives out of a total of six tested is regarded as an
unexpected result.
NOTE: At LOD50, 50 % of the
samples are found positive. In theory [taking into account that the method used
is capable of detecting a single microorganism in a sample and assuming a
homogeneous (Poisson) distribution between the samples] the average
contamination level of the samples is expected to be 0,7 microorganisms per
sample in order to reach the LOD50. In practice, the ideal homogeneous
distribution is often not reached and relatively more of the samples do not
contain any (viable) microorganism than could be expected for a true
homogeneous distribution. In order to obtain 50 % of positive samples, an
increase in the average level of contamination is required, based on experience
with the material used in the study.
8.4.3. Scheme comparisons of laboratory
performance
To compare the performance of one laboratory
against other participating laboratories, the scheme organizers may calculate
the numbers (or percentages) of positives for the levels contaminated with the
target organism found per test sample by each laboratory (reported by
laboratory code). An example of such data is given in Figure 2. In this study 28
laboratories participated (indicated as lab codes on the NL-axis
of Figure 2). Each laboratory analysed 22 chicken faeces samples, artificially
contaminated with reference materials containing two different Salmonella
serovars at four different contamination levels (varying from 10 to 500 cfu per
sample). In Figure 2 the results are summarized for the sample with the low
level of contamination (n = 14). The number of positives found in the
participating laboratories varied from 0 to 11 samples (n+ -axis
in Figure 2).
In addition to this more descriptive way of
presenting the results, it is possible to calculate specificity rates,
sensitivity rates and accuracy rates per level of contamination of the samples
(ISO 16140[4]). These rates may be calculated for each laboratory
and for the results from all laboratories.

KEY:
NL lab code
n+ No. of positives
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The specificity rate, rSP,
is given by:

where:
n_ is the number of negative results found;
E(n− tot) is the total number
of expected negative samples.
The sensitivity rate, rSE, is
given by:

where:
n+ is the number of positive
results found;
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
The accuracy rate, rAC, is
given by:

where ntot is
the total number of samples.
This assessment is only meaningful if linked
with the number of target organisms present.
ANNEX A
(informative)
EXAMPLE
OF DETAILS TO BE INCLUDED IN A PT SCHEME PLAN
PT scheme plan -
Summary of scheme
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Food examinations scheme
Scheme type:
Food microbiology
Aims:
To provide external quality assessment
samples for general routine examinations undertaken by food microbiology
laboratories
Criteria for selection of participants:
Food microbiology laboratories with
laboratory facilities adequate for dealing with pathogenic microorganisms of
risk categories 1 and 2
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Food microbiology laboratories in the
private and public sectors
Legislation:
EU Regulation 882/2004 concerning the official
control of foodstuffs
Sample type:
Freeze-dried microorganisms in evacuated
glass vials
Examinations:
Presence/absence:
Campylobacteria spp.
Escherichia coli O157
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Enumerations:
Aerobic colony count
Bacillus cerius
Clostridium perfringens
Coliforms
Enterobacteriaceae
Escherichia coli
Listeria monocytogenes
Coagulase positive staphylococci
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Realistic microflora simulating that of
real foods and providing a realistic challenge to routine food microbiology
procedures
Target number of participants:
More than 200
Number of distributions per year:
Six (6)
Number of samples per distribution:
Two (2)
External subcontractors:
None
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Named individuals
Quality control testing of sample batch:
Name of provider laboratory and standard
methods used. 25 samples from every batch examined for all tests specified
Statistical methods:
Consensus median (enumerations)
Percentiles to identify outliers
Allocation of scores:
Yes
Criteria for scores:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Continuous performance assessment:
Yes
Criteria for identifying “poor performers”:
Less than 70 % maximum possible score over
six distributions
Proactive approach by organizers to “poor
performance”:
Yes
Method assessment:
Yes - for indicators only
Scheme co-ordinator
or deputy to sign/date
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Date
Promotional literature authorized:
Date
Costing approved:
Date
Accreditation dated:
Date
Other comments:
Review at steering group meeting
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
ANNEX B
(informative)
METHODS
OF TESTING FOR VARIATION BETWEEN PORTIONS OF TEST MATERIALS
B.1. T1 − T2 test
This test is recommended for cases where low
numbers of organisms are present in portions of test materials (TMs) at levels
up to 35 to 40 cfu per plate or where no “target standard deviation” can be
assigned for assessing sufficient homogeneity.
The variation between analytical portions
from one (reconstituted) unit of TM, T1, and that between analytical
portions from different (reconstituted) units of one batch of TM, T2,
is tested in different ways. Details can be found in Reference [12], but a
summary is given here.
For the determination of the variation
between analytical portions of one (reconstituted) unit of an TM (replicate
testing), the T1 test statistic is applied:
(B.1)
where:
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
zi+ is the sum of numbers of cfus
in all analytical portions of unit i
(B.2)
J is the number of analytical portions per
unit.
For the determination of the variation
between analytical portions from different (reconstituted) units of one batch
of TM, the T2 test statistic is applied:
(B.3)
where:
z++ is the sum of numbers of cfus
in all analytical portions of the tested units of one batch of TMs

I is the number of units tested.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
For the variation between units of one batch
of TM, the Poisson distribution is the theoretical smallest possible variation
that could be achieved. However, overdispersion is expected and
is mostly larger than 1 (Reference
[12]). An acceptable variation between units of a batch of TM is
≤ 2.
EXAMPLE
Given the following data:
Unit:
(duplicate) counts:
1
z11 = 45 z12 =
49
2
z21 = 33 z22 =
42
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
z31 = 40 z32 =
42
I = 3 (three units)
J = 2 (two replicates)
z1j/J = (45+49)/2 = 94/2 = 47
z2j/J = (33+42)/2 = 75/2 = 37,5
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.

T1 should follow a
-distribution with I(J − 1) = 3 × (2 −
1) = 3 degrees of freedom.
Tested two-sided at the 95 % confidence
level, the lower and upper limit for this distribution are, with 3 degrees of
freedom, 0,22 and 9,3, respectively. The calculated T1 value
(1,298) follows these criteria.

Accepted variation for the batch is: T2
/(I - 1) ≤ 2.
Here T2 /(I - 1) = 2,206/(3
- 1) = 1,103 and thus follows the criteria for acceptability of the batch.
B.2. Test for sufficient homogeneity
This test is recommended for cases where
larger numbers of organisms (more than 35 to 40 cfu per plate) are present in
portions of the test material and a target standard deviation, sp, that describes the
performance expected of PT scheme participants is available. It is based on the
“sufficient homogeneity” test of Reference [11].
Given a set of test material portions
analysed in duplicate with results expressed in log units, the test is passed
if the between-portion variance, S2sam, satisfies
Condition (B.5):
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
where S2an is the
analytical variance. This test is less likely than that of T1 − T2
to lead to the rejection of a test material that is not perfectly homogenous
(analytical results follow the Poisson distribution), but is sufficiently
homogenous to be used in a proficiency test round with a target standard
deviation of sp. This is because
materials are accepted unless it is shown with high confidence (95 %) that the
fitness for purpose criterion of ssam > 0,3 sp, where ssam is the standard
deviation of which ssam is an estimate.
EXAMPLE:
Given quantitative results from the duplicate
analysis of 10 test material portions, calculate the difference (D) and sum (S)
and the square of D (D2) of the decimal logarithm of each set of
results (Table 1).
Portion
Analysis 1
Analysis 2
Log analysis 1
Log analysis 2
D
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
D2
1
35
51
1,5441
1,7076
-0,1635
3,2516
0,026733
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
52
46
1,7160
1,6628
0,0532
3,3788
0,002835
3
35
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
1,5441
1,5185
0,0256
3,0626
0,000653
4
53
38
1,7243
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,1445
3,3041
0,020878
5
30
40
1,4771
1,6021
-0,1249
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,015610
6
33
30
1,5185
1,4771
0,0414
2,9956
0,001713
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
41
60
1,6128
1,7782
-0,1654
3,3909
0,027346
8
35
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
1,5441
1,7404
-0,1963
3,2844
0,038532
9
68
67
1,8325
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,0064
3,6586
0,000041
10
52
60
1,7160
1,7782
-0,0621
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
0,003362
Calculate the sum of D2 and
divide it by twice the number of portions. This is the mean sum of squares of
within sample variation,
In this case,
=
0,138 2/20 = 0,00691.
Calculate the variance of
and divide it by two. This is equal to
.
In this case,
=
0,04224/2 = 0,02112
Then S2an =
and S2sam = (
-
)/2
In this case, S2an =
0,00691 and S2sam = (0,02112 - 0,00691)/2 = 0,007104
Values for F1 and F2
depend on the number of portions examined in the test. For 10 portions F1
= 1,88 and F2 = 1,01 (values for other numbers of portions
may be found in Reference [11]).
If the target standard deviation to be
applied to the proficiency test results, σp, is equal to 0,25 log10
units then
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
This value is larger than S2sam
(0,007104). Hence Condition (B.5) is satisfied and the test material is
sufficiently homogenous.
ANNEX C
(informative)
A
PRACTICAL METHOD TO ASSESS LONG-TERM PERFORMANCE OF PARTICIPANTS IN PT SCHEMES
USING ENUMERATION METHODS
C.1. Preparing data for analysis
Four columns of a spreadsheet are required to
analyse enumeration results:
a) the count as reported by the participant (nR-count);
b) the count to be used for charts and
histograms and/or for scoring (nS-count);
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
d) a comment column (text field) (nC-count)
for recording any notes about participants' results.
The nR-count should
normally be entered as a numerical field (e.g. 1 100) or a scientific number
(e.g. 1.1e3). If the result is reported as a censored value, then a text field
shall be used and a “less than” or “greater than” symbol entered in front of
the number (e.g. < 10, > 1 100, etc.).
The nR-count may also be
entered as other text such as NE (not examined), ND (not detected), and UA
(unassessable). Entries shall be made for all fields in the column.
If the nR-count cannot be
analysed (e.g. entered as NE) then there is no further analysis or allocation
of score.
If the nR-count cannot be
included in the statistical analysis, e.g. the result was reported as a
censored value, but a score is to be allocated and/or the result is to be
plotted, then a numerical value shall be assigned to the ns-count.
This value shall ensure that the correct
score is allocated and that the result is plotted correctly. For example,
scores may not be allocated to results reported as “not detected” but it may be
helpful to include those results on a chart. In this case, the ns-count
may be entered as −99 and the nA-count should be left blank.
If a reported result is to be allocated, a
score is to be included in the statistical calculations; then the nA-count
= ns-count.
C.2. Dealing with censored data
All low censored results are analysed in one
of the following ways.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
b) All low censored values and results of
zero or “not detected” are assigned an ns-count of 0,21) and
included in the analysis because the level of the target organism or group was
relatively low and those results may have arisen by chance (ns-count
= nA -count = 0,2). There are exceptions, such as when the
reported low censored value (<x) shows an inappropriate level of detection
and the value of x is actually higher than the median (on initial
calculation). In these exceptional cases the nA-count should
be left blank.
c) Scores are not allocated to low censored
values. In general, scores should be allocated unless there is a
microbiological reason for not doing so. In this case, the nA-count
and ns-count fields are left blank.
High censored values are entered as 1,0 log10
above the maximum reported score. If, for any reason, the results are to
be excluded from the charts and analysis, and no scores are to be allocated,
then the ns-count and nA-count fields
should be left blank. If a result is reported as >x where the x-value is
less than the median then the nA-count field should be left
blank.
C.3. Plotting results
Plots are based on the nS-count
values. There are two main types of plots used for PT scheme results:
histograms (or bar charts) and scatter plots. Points to be considered when
choosing which type of plot to use include the type of examination (MPN, colony
counts, etc.) and the number of participants undertaking the examination.
Histograms are produced from the decimal
logarithm value of the nR-count grouped as follows:
< 0, (0 to 0,05), (0,05 to 0,1), (0,1 to
0,15), (max. log10 nR-count), (max. log10 nR)-count + 0,05]
Any non-numeric or special cases to be
plotted (e.g. −99) should be given their own bar.
Alternatively, the histogram may be produced
based on the decimal logarithm value of the nR-count rounded
to the nearest 0,05. In this case the groupings are as follows.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
In this case, the bar 0,1 includes all
results from 0,075 to 0,124.
Scatter plots can be produced using the nS-count
values directly then converting the y-axis to a decimal logarithmic scale. Non-numeric
data are excluded from scatter plots
Bin-sizes (groups) of 0,05 log10
are normally used for the histograms so the ranges for allocation of scores are
also rounded to 0,05 log10.
C.4. Allocation of scores
Where PT scheme results are assessed using
scores, the criteria for allocation of scores should be listed in scheme
protocols or reports. The score for the enumeration result is entered in a
score field (nC-score); this may be manually amended where necessary.
Note that the final score for a result may be
reached from the nC-score, but other factors may also require consideration.
ANNEX D
(informative)
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Safety data sheet for food PT scheme samples
- Freeze-dried
Effective date: dd-mm-yy
Review date: dd-mm-yy
Issued to: All scheme
participants
Identification of the product and the
establishment
Product: Simulated food
samples for general microbiological examinations
Establishment: Full address and
contact details for scheme organizer
Composition or information on ingredients
Glass vials of freeze-dried material
containing a mixture of bacteria of hazard group 2 as defined by national and
international legislation. A hazard group 2 organism may cause human disease
and may be a hazard to laboratory workers, but is unlikely to spread to the
community.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Physicochemical hazard: Not
applicable
Health hazard: Minimal risk of infection,
provided good laboratory practice is observed
Environmental hazard: Not applicable
First aid measures
If accidental contact with material occurs,
laboratory staff shall follow local first aid procedures that are normally
applied following exposure to an equivalent food sample. Following exposure to
the material, medical advice shall be sought.
Fire fighting measures
Not applicable
Accidental release measures
Cover the area with absorbent material and
flood with a suitable disinfectant. The area shall be left undisturbed for 30
min before the spill is mopped up with an excess of absorbent material. Wear
appropriate personal protective equipment.
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Store at room temperature in the dark.
Samples shall be processed in a laboratory environment which, as defined by national
regulations or guidelines, is suitable for the practice of microbiology. Staff
handling the material should have been trained in the handling of infectious
biological material. The material should be treated with the same degree of
care as would be exercised with equivalent food samples. Hand-tomouth contact
should be avoided while working with the samples and normal hand-washing
procedures relating to the handling of routine samples shall also be observed
with PT samples.
Exposure controls and personal protection
Use good laboratory practice and wear
appropriate laboratory coats, gloves and eye protection. Removal of vials from
packaging and reconstitution should be carried out in an exhaust protective
cabinet.
Physical and chemical properties
Inert odourless dry material.
Stability and reactivity
Storage is unlikely to increase or decrease
the risks of infection associated with handling the material.
Toxicological information
Not applicable
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
Not applicable
Disposal considerations
The used material shall be disposed of using
an autoclave as for food products containing infectious microorganisms and in
accordance with all local and national regulations.
Transport information
Refer to national and international
regulations for transport of bacteria in hazard group 2 (biological substance,
category B; UN3373).
Regulatory information
EC Biological agent, hazard category/risk
group 2
CAUTION - This safety data sheet does not
constitute the user's own assessments of workplace risk as required by health
and safety legislation.
Other information
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
For further safety information concerning
this product, participants are advised to read the instruction sheet
accompanying the samples.
BIBLIOGRAPHY
[1] TCVN 6910-2 (ISO 5725-2) Accuracy
(trueness and precision) of measurement methods and results - Part 2: Basic
method for the determination of repeatability and reproducibility of a standard
measurement method.
[2] TCVN 6910-4 (ISO 5725-4) Accuracy
(trueness and precision) of measurement methods and results - Part 4: Basic
methods for the determination of the trueness of a standard measurement method.
[3] TCVN 8128-1:2009 (ISO/TS 11133-1:2009) Microbiology
of food and animal feeding stuffs — Guidelines on preparation and production of
culture media - Part 1: General guidelines on quality assurance for the
preparation of culture media in the laboratory
[4] ISO 16140 Microbiology of food and
animal feeding stuffs - Protocol for the validation of alternative methods
[5] TCVN ISO/IEC 17025 General
requirements for the competence of testing and calibration laboratories
[6] ISO/TS 19036 Microbiology of food and
animal feeding stuffs - Guidelines for the estimation of measurement
uncertainty for quantitative determinations
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.
[8] ILAC-G13, ILAC Guidelines for the requirements
for the competence of providers of proficiency schemes. Available
(2010-02-10) at: http://www.ilac.org/documents/ILAC G13 08 2007.pdf
[9] Thompson, M., ELLISON, L.R,. WOOR, R. for
IUPAC. The international harmonized protocol for the proficiency testing of
analytical chemistry laboratories. Pure Appl. Chem.. 2006, 78,
pp. 145- 196. Available (2010-10-08) at: http://iupac.org/publications/pac/2006/pdf/7801x0145.pdf
[10] AUGUSTIN, J.C., CARLIER, V. Lessons from
the organization of a proficiency testing program in food microbiology by
interlaboratory comparison: analytical methods in use, impact of methods on
bacterial counts and measurement uncertainty of bacterial counts. Food
Microbiol. 2006, 23, pp. 1-138
[11] FEARN, T., THOMPSON, M. A new test for
“sufficient homogeneity”. Analyst 2001, 126, pp, 1414-1417
[12] HEISTERKAMP, S.H., HOEKSTRA, J.A., VAN
STRIJP-LOCKEFEER, N.G.W.M,. VAVELAAR, A.H,. MOOIJMAN, K.A, IN’T VELD, P.H.,
S.H.W., MAIER, E.A.; GRIEPINK.B. Statistical analysis of certification trials
for microbioligcal reference materials. Luxembourg: commission of the
European communities, 1993. 41 p. (Report EUR 15008 EN.)
[13] JARVIS, B. Sampling for microbiological
analysis. In: LUND, B.M., BAIR-PACKER, A C., GOULD, G.W., editors. The
microbiological safety and quality of food, Vol. 2, pp.1691-1734.
Gaithersburg, MD: Aspen, 2000
[14] JARVIS. B. Statistical aspects of the
microbiological examination of foods, 2nd edition. Amsterdam:
Academic Press, 2008. 306 p.
[15] JARVIS, B., HEDDGES.A.J., CORRY, J.E.L.
Assessment of measurement uncertainty for quantitative methods of analysis:
comparative assessment of the precision (uncertainty) of bacterial colony
counts. Int. J. Food Microbiol. 2007, 116, pp. 44-51
[16] JARVIS.B., CORRY, J.E.R., HEDGES, A.J. Estimates
of measurement uncertainty from proficiency testing schemes, internal
laboratory quality monitoring and during routine enforcement examination fo
foods. J. Appl. Microbiol. 2007, 103, pp. 462-467
...
...
...
Hãy đăng nhập hoặc đăng ký Thành viên
Pro tại đây để xem toàn bộ văn bản tiếng Anh.